HEAT STRESS MONITOR ( WBGT ) SK-150GT PT OCTA PRIMA LESTARI
Perhatian!
Spesifikasi HEAT STRESS MONITOR ( WBGT ) SK-150GT PT OCTA PRIMA LESTARI
The SK-150GT is a handy type instrument that can measure temperature, humidity and globe temperature to approximate the WBGT index ( * ) .
This can be used as a guide for preventing heat related illness and estimating heat stress of working and activity related environments.
WBGT Index :
The process for determining the WBGT ( Wet Bulb Globe Temperature) is described in ISO 7243 / JIS Z8504, Hot Environments
- Estimation of the Heat Stress on Working Man, based on the WBGT Index. The WBGT index is a sensory temperature which can be calculated from wet bulb temperature, globe temperature ( radiation heat) and dry bulb temperature. It is necessary to set the measuring ambient before starting since different calculation is needed depending on its measuring ambient.
( 1) Indoor ( IN) : when measuring indoor or outdoor away from direct sun light.
WBGT Index( ° C) = 0.7 x Wet Bulb Temperature( ° C) + 0.3 Globe x Temperature( ° C)
( 2) Outdoor ( OUT) : when measuring outdoor, under direct sun light.
WBGT Index( ° C) = 0.7 x Wet Bulb Temperature( ° C) + 0.2 x Globe Temperature( ° C) + 0.1 × Dry Bulb Temperature( ° C)
Heat Stress :
Heat stress is a condition that temperature regulation function decreases by high temperature ambient. This could cause a heat related illness and even ultimately death. The risk can be reduced by observing the surrounding ambient and by being aware of precautions of heat stock such as overheating and dehydration.
FEATURES :
Alarm: The alarm is triggered when the measured WBGT value reaches the set value.
HOLD function, MAX/ MIN function
Measuring ambient of WBGT index can be set.
Auto power-off ( in 20 min.)
Carrying case for damage prevention and storing of the instrument.
SPECIFICATIONS
Measuring range Air Temperature ( TA) 0.0 to 50.0° C
Globe Temperature ( TG) 0.0 to 80.0° C
WBGT Index 0.0 to 50.0° C
Humidity 10.0 to 95.0% RH
Accuracy Air Temperature ( TA) ± 0.6° C
Globe Temperature ( TG) ± 2° C
WBGT Index ± 2° C
Humidity ± 3% RH ( 20 to 90% rh at20 to 30° C)
± 5% rh ( other than above)
Resolution Air Temperature ( TA) 0.1° C
Globe Temperature ( TG) 0.1° C
WBGT Index 0.1° C
Humidity 0.1% RH
Sampling time approx. 2 seconds
WBGT alarm Setting range 20.0 to 49.9 ( no cancelation)
Functions Change-over switch for WBGT Index of indoor and outdoor
Auto power-off in approx. 20 min.
MAX/ MIN
HOLD
Operation ambient 0 to 50° C, below 95% RH ( no condensing)
Storage ambient -10 to 50° C ( no condensing)
Power requirement 9VDC Battery 6F22 ( 006P) x 1
Battery life approx. 600 hours in continuous use ( without use of alarm)
Dimensions Main Unit : ( W) 71 x ( H) 27.5 x ( D) 30 mm ( without globe)
Globe : Ø 60 mm
Weight approx. 240 g ( with battery)
Standard accessories 9VDC Battery 6F22 ( 006P) x 1
Carrying case
PT OCTA PRIMA LESTARI
Rukan Mitra Matraman Blok B-12
Jl. Matraman Raya No. 148 – Jakarta 13150
Tlp : 85902391 – 85918154- 8512595 – 85918155
Fax : 85918178
Web : www.octaprimalestari.com
OCTA PRIMA LESTARI
Alamat : Rukan Mitra Matraman Blok B-12 , Jl. Matraman Raya No. 148 - Jakarta
Tlp : 85918154 - 85918155 - 85902391 - 8512595
E-mail : sales@octaprimalestari.com
Web : www.octa primalestari.com
Perusahaan kami bergerak dalam bidang peralatan laboratorium untuk analysa pabrik sbb :
- ALAT UJI TEST KEBOCORAN PACKAGING ( VACUUM DESSICATOR ) LOKAL
- ANALYTICAL BALANCE OHAUS
- BROOKFIELD VISCOMETER
- BOD INCUBATOR LOKAL
- CONDUCTIVITY METER
- DO METER
- FLOWATCH FLOW METER
- GAST VACUUM PUMP MODEL DOA P-504 BN
- GRAB SAMPLER/ EKMAN GRAB LOKAL
- HIGH VOLUME AIR SAMPLER STAPLEX
- LAMINAR AIR FLOW LOKAL
- LEMARI ASAM LOKAL
- LUX METER PCE
- PH METER ADWA INSTRUMENTS
- SOUND LEVEL PCE
- THERMOHYGROGRAPH PCE
- WEATHER STATION METER PCE
dll