Voltage Test |
Voltage Range |
12 - 690V AC/DC |
LED |
Nominal Voltage |
12/24/50/120/230/400/690V
AC (16 - 400Hz), DC (±) |
Tolerance
(Threshold Voltage) |
Light on at more than: |
|
7±3V (12V LED) |
|
18±3V (24V LED) |
|
37.5±4V (50V LED) |
|
75%±5% of nominal voltage |
|
(120/230/400/690V LED) |
|
Response Time |
<0.6s at 100% of each nominal voltage |
LCD |
Range/Resolution
(Auto-range) |
300V AC/DC (6.0 - 299.9)/0.1V
690V AC (270 - 759)/1V
690V DC (270 - 710)/1V |
Accuracy (23±5ºC) |
±1.5V (7 - 100V)
±1%±5dgt (100 - 690V)
AC (16 - 400Hz), DC(±) |
Over limit Indication |
"OL" |
Response Time |
Approx. 1s at 90% - 110% of each voltage |
Peak Current |
ls<3.5mA (at 690V) |
Measurement Duty |
30s ON (operation time)
240s OFF (recovery time) |
Single-pole Phase Test |
Voltage Range |
100 - 690V AC (50/60Hz) |
Phase Rotation Test |
System |
Three-phase 4-wire system
200 - 690V phase-to-phase AC (50/60Hz) |
Phase Range |
120±5 degree |
Continuity Test |
Detection Range |
0 - 400kΩ+50% |
Test Current |
Approx. 1.5μA (battery 3V, 0Ω) |
General |
Operating temperature and humidity ranges |
-15 - 55ºC, max 85% RH (No Condensation) |
Storage temperature and humidity ranges |
-20 - 60ºC, max 85% RH (No Condensation) |
Applicable Standards |
IEC 61243-3, IEC 61010-1, IEC 61010-031, IEC 61557-7
CAT Ⅲ 690V / CAT Ⅳ 600V Pollution degree 2 |
IP Rating |
IP65 (IEC 60529) |
Power Source |
LR03 (AAA) 1.5V×2 |
Dimensions |
246(L)×64(W)×26(D) mm |
Weight |
190g (including batteries) |
Accessories |
LR03 (AAA) 1.5V×2, KTA01 (4mm metal tips [2pcs/set]),
KTA02 (4mm rubber caps [2pcs/set]), Instruction manual |